Context
On Ubuntu, license activation in Thermo Scientific License Manager (LM) may fail with the following error:
FlexNet initialization error: 2
This issue can occur when Thermo Scientific License Manager is installed using dpkg -i without the required libfuse2 dependency. In this case, the application files may be unpacked, but the installation may stop before completion, preventing the FlexNet Licensing Service from being installed correctly.
Process
1. Remove the incomplete Thermo Scientific License Manager installation:
sudo apt remove thermo-scientific-license-manager
2. Install the required dependency:
sudo apt install libfuse2
3. Reinstall Thermo Scientific License Manager:
sudo apt install ./thermo-scientific-license-manager_<version>_amd64.deb
4. Retry the license activation.
5. If the error persists, manually reinstall the FlexNet Licensing Service:
sudo /opt/thermo-scientific-license-manager/external_binaries/flexnet/bin/licensingService/install_fnp.sh /opt/thermo-scientific-license-manager/external_binaries/flexnet/bin/licensingService/FNPLicensingService
Make sure the command completes without errors, then retry the license activation.
6. If the error still persists, make sure that the permissions allow write access to the FlexNet Trusted Storage directories:
sudo chmod -R a+rwX /usr/local/share/macrovision
Retry the license activation.
Note: On Ubuntu, use:
sudo apt install ./<package>.deb
instead of:
sudo dpkg -i <package>.deb
Using apt resolves required dependencies, such as libfuse2, before completing the installation.
Do not delete or manually recreate the FlexNet Trusted Storage files.
If the issue persists after completing the steps above, contact our Customer Support for further assistance.